Yazar "Ozturk, Caglar" seçeneğine göre listele
Listeleniyor 1 - 1 / 1
Sayfa Başına Sonuç
Sıralama seçenekleri
Öğe Numerical investigation of volute tongue design on hemodynamic characteristics and hemolysis of the centrifugal blood pump(Springer Int Publ Ag, 2021) Aka, Ibrahim Basar; Ozturk, Caglar; Lazoglu, IsmailIn the design of rotary blood pumps, the optimization of design parameters plays an essential role in enhancing the hydrodynamic performance and hemocompatibility. This study investigates the influence of the volute tongue angle as a volute geometric parameter on the hemodynamic characteristics of a blood pump. A numerical investigation on five different versions of volute designs is carried out by utilizing a computational fluid dynamics (CFD) software ANSYS-FLUENT. The effect of volute tongue angle is evaluated regarding the hydrodynamic performance, circumferential pressure distribution, the radial force, and the blood damage potential. A series of volute configurations are constructed with a fixed radial gap (5%), but varying tongue angles ranging from 10 to 50 degrees. The relative hemolysis is assessed with the Eulerian based empirical power-law blood damage model. The pressure-flow rate characteristics of the volute designs at a range of rotational speeds are obtained from the experimental measurements by using the blood analog fluid. The results indicate an inverse relationship between hydraulic performance and the tongue angle; at higher tongue angles, a decrease in performance was observed. However, a higher tongue angle improves the net radial force acting on the impeller. The pump achieves the optimized performance at 20 degrees of the tongue angle with the relatively high hydrodynamic performance and minor blood damage risk.