Radioactive source calibration test of the CMS Hadron Endcap Calorimeter test wedge with Phase I upgrade electronics

dc.WoS.categoriesInstruments & Instrumentationen_US
dc.authorid0000-0002-9007-8260en_US
dc.contributor.authorYetkin, Elif Aslı
dc.date.accessioned2020-11-16T09:28:06Z
dc.date.available2020-11-16T09:28:06Z
dc.date.issued2017-12
dc.departmentFakülteler, Mühendislik ve Doğa Bilimleri Fakültesi, Enerji Sistemleri Mühendisliği Bölümüen_US
dc.description.abstractThe Phase I upgrade of the CMS Hadron Endcap Calorimeters consists of new photodetectors (Silicon Photomultipliers in place of Hybrid Photo-Diodes) and front-end electronics. The upgrade will eliminate the noise and the calibration drift of the Hybrid Photo-Diodes and enable the mitigation of the radiation damage of the scintillators and the wavelength shifting fibers with a larger spectral acceptance of the Silicon Photomultipliers. The upgrade also includes increased longitudinal segmentation of the calorimeter readout, which allows pile-up mitigation and recalibration due to depth-dependent radiation damage. As a realistic operational test, the responses of the Hadron Endcap Calorimeter wedges were calibrated with a Co-60 radioactive source with upgrade electronics. The test successfully established the procedure for future source calibrations of the Hadron Endcap Calorimeters. Here we describe the instrumentation details and the operational experiences related to the sourcing test.en_US
dc.fullTextLevelFull Texten_US
dc.identifier.doi10.1088/1748-0221/12/12/P12034
dc.identifier.scopus2-s2.0-85039773146en_US
dc.identifier.scopusqualityQ3en_US
dc.identifier.urihttps://hdl.handle.net/11411/2656
dc.identifier.urihttps://doi.org/10.1088/1748-0221/12/12/P12034
dc.identifier.wosWOS:000419046800001en_US
dc.identifier.wosqualityQ3en_US
dc.indekslendigikaynakWeb of Scienceen_US
dc.indekslendigikaynakScopusen_US
dc.language.isoenen_US
dc.nationalInternationalen_US
dc.numberofauthors50+en_US
dc.publisherIOP PUBLISHING LTDen_US
dc.relation.ispartofJOURNAL OF INSTRUMENTATIONen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectDetector alignment and calibration methods (lasers, sources, particle-beams)en_US
dc.subjectCalorimeter methods; Calorimetersen_US
dc.titleRadioactive source calibration test of the CMS Hadron Endcap Calorimeter test wedge with Phase I upgrade electronics
dc.typeArticle
dc.volume12en_US

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
2017 WOS Yetkin.pdf
Boyut:
2.81 MB
Biçim:
Adobe Portable Document Format
Açıklama:
Lisans paketi
Listeleniyor 1 - 1 / 1
Küçük Resim Yok
İsim:
license.txt
Boyut:
1.71 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: