dc.description28th European Conference on Solid-State Transducers (EUROSENSORS) -- SEP 07-10, 2014 -- Brescia, ITALYen_US
dc.description.abstractA fully optical active catheter tracking system compatible with 3T MRI environment is presented. It replaces conducting cables with optical fibers to reduce RF-induced heating problem. Proposed system consists of a MEMS-based microstructure array and an IC driving it. The IC houses an RF receiver block and an optoelectrical power supply. A prototype IC was fabricated in UMC 0.18 mu m CMOS process. Measurements indicate that the supply unit is able to provide 2.18mA at 1.2V supply, when a laser beam of 80mW power at its source is applied to IC. The system is operational at laser source power levels above 40mW. (C) 2014 The Authors. Published by Elsevier Ltd.en_US
dc.description.sponsorshipNatl Res Council,AISEMen_US
dc.identifier.doi10.1016/j.proeng.2014.11.725
dc.identifier.endpage1464en_US
dc.identifier.issn1877-7058
dc.identifier.scopus2-s2.0-84923345450en_US
dc.identifier.scopusqualityN/Aen_US
dc.identifier.startpage1461en_US
dc.identifier.urihttps://doi.org/10.1016/j.proeng.2014.11.725
dc.identifier.urihttps://hdl.handle.net/11411/8842
dc.identifier.volume87en_US
dc.identifier.wosWOS:000369531700353en_US
dc.identifier.wosqualityN/Aen_US
dc.indekslendigikaynakWeb of Scienceen_US
dc.indekslendigikaynakScopusen_US
dc.language.isoenen_US
dc.publisherElsevier Science Bven_US
dc.relation.ispartof28th European Conference on Solid-State Transducers (Eurosensors 2014)en_US
dc.relation.publicationcategoryKonferans Öğesi - Uluslararası - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectMrıen_US
dc.subjectOptical Poweringen_US
dc.subjectCmos Photodiodeen_US
dc.subjectElectromagnetic Actuationen_US
dc.subjectOptical Memsen_US
dc.titleAn Optoelectrical, Standard CMOS-Based Active Catheter Tracking System for MRIen_US
dc.typeConference Objecten_US

Dosyalar