Implementation of k0-INAA standardisation at ITU TRIGA Mark II research reactor, Turkey based on k0-IAEA software
No Thumbnail Available
Date
2016
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Pergamon-Elsevier Science Ltd
Access Rights
info:eu-repo/semantics/closedAccess
Abstract
The purpose of this study is to test the applicability of k(0)-INAA method at the Istanbul Technical University TRIGA Mark II research reactor. The neutron spectrum parameters such as epithermal neutron flux distribution parameter (a), thermal to epithermal neutron flux ratio (f) and thermal neutron flux (phi th) were determined at the central irradiation channel of the ITU TRIGA Mark II research reactor using bare triple-monitor method. HPGe detector calibrations and calculations were carried out by k(0)-IAEA software. The alpha, f and phi th values were calculated to be -0.009, 15.4 and 7.92.10(12) cm(-2) s(-1), respectively. NIST SRM 1633b coal fly ash and intercomparison samples consisting of clay and sandy soil samples were used to evaluate the validity of the method. For selected elements, the statistical evaluation of the analysis results was carried out by z-score test. A good agreement between certified/reported and experimental values was obtained. (C) 2015 Elsevier Ltd. All rights reserved.
Description
Keywords
K(0)-Inaa, K(0)-Iaea, Neutron Spectrum Parameters, Trıga, Neutron-Activation Analysis, Validation, Alpha, Ratio
Journal or Series
Radiation Physics and Chemistry
WoS Q Value
Q1
Scopus Q Value
Q2
Volume
119